FLUOROSCOPY PERFORMANCE - ABC [Latest Technology]

This is a function of Fluoroscopy, which controls the overall performance of the fluoroscopic image by automatically adjusting kVp, mA or both. By this, the image brightness is kept fixed on the monitor. It is a feed back circuit that measures the light intensity or video camera signal of the output screen.
FLUOROSCOPY

Image A photomultiplier tube or a photocathode is used to monitor the light output of the intensifier tube. When the corresponding brightness changes, it is sent to the generator for adjustment. The generator controls the exposure rate by changing kVp, mA or both. The brightness of the central area of ​​the output screen is measured for this adjustment.

Brightness can be adjusted with both kVp, mA, which affects both contrast and parient dose. Three methods of adjustment are as follows -

  • Change of kV at constant mA
  • Change of mA at constant kV
  • Change of both kV and mA

When the II tube moves from the thin part of the patient to the thick part, then increasing the kV reduces the patient dose and contrast. In addition, when mA is increased, contrast is obtained better and patient dose is also higher.
input dose rate

Generally mA is first adjusted to adjust the input dose rate of the II tube. The kV is adjusted to the input dose rate when mA reaches its maximum limit.

⇒ For More Job Updates: Click Here
⇒ Join Our WhatsApp Group: Click Here


Important: Please always Check and Confirm the above details with the official website and Advertisement / Notification.
0 C "FLUOROSCOPY PERFORMANCE - ABC [Latest Technology] "

© Copyright 2020 at www.socioeducation.in

Disclaimer :
The Examination Results / Marks published in this Website is only for the immediate Information to the Examinees an does not to be a constitute to be a Legal Document. While all efforts have been made to make the Information available on this Website as Authentic as possible. We are not responsible for any Inadvertent Error that may have crept in the Examination Results / Marks being published in this Website nad for any loss to anybody or anything caused by any Shortcoming, Defect or Inaccuracy of the Information on this Website.
Back To Top